Survey Meter DGM-1500 Turva
|
|
"Product PDF" |
|
Spicifications :
|
RADIATION DETECTED |
Gamma and X-ray |
MEASUREMENT UNITS |
microSv/h (μSv/h) |
LINEAR RANGE |
Dose rate 0.01 - 100 000 microSv/h Dose 0.001 - 1000 milliSv |
SENSITIVITY |
150 cpm/microSv/h |
ACCURACY |
± 5 % of the reading when irradiated by Cs-137 at the calibration point at 20 deg. °C. |
LINEARITY |
± 10 % at 20 °C |
RESPONSE TIMES |
Fast measurement mode 2.5 sec.
Automatic measurement mode 3 min. at ambient radiation levels, at increased radiation levels (> 100 microSv/h) 5 sec. Accuracy measurement mode, at low level radiation, integration time long to 3 hours as far as, used in gamma radiation measuring of foodstuff, for example. |
DETECTOR TYPE |
Ambient dose equivalent Hp(10) -energy compensated GM-tube, quenching type Neon/Halogen |
ENERGY DEPENDENCE |
± 30 %, 30 keV - 1.25 Mev |
PHYSICAL CHARACTERISTICS |
Display type 4 -digit LCD-display 44 x 18 mm. Big dose rate showed like this 30E3 (30 000 microSv/h) |
OPERATING TEMPERATURE RANGE |
-30°C to +55°C |
OVERALL DIMENSIONS |
L x W x H1/H2, 14.7cm x 9.1cm x 5.2/3.2cm |
WEIGHT |
300 g with battery |
AUDIO SPEAKER |
Piezoelectric speaker |
ALARMS |
Adjustable dose rate and dose alarm level |
BATTERY |
(1) 6LR61, 9 V battery, battery life 300 h (background level), low-battery indication |
MAIN ADAPTER |
9 VDC reg. / 2.5 W (3,5 mm plug) diode protected against wrong polarity
Standard configuration incl. battery, calibration certification and user manual. |
|
|
|
|
|
|
|
|
|
|